LAUDA LCK 2907 Alpha RA 8

LAUDA LCK 2907 Alpha RA 8 jahutustsirkulaatori kasutusjuhend

Mudel: LCK 2907 Alpha RA 8

Kaubamärk: LAUDA

1. Sissejuhatus

This manual provides essential information for the safe and efficient operation, setup, and maintenance of your LAUDA LCK 2907 Alpha RA 8 Cooling Circulator. Please read this manual thoroughly before operating the device to ensure proper function and to prevent damage or injury.

The LAUDA LCK 2907 Alpha RA 8 is designed for precise temperature control in various laboratory and industrial applications, offering cooling performance across a wide temperature range.

2. Ohutusjuhised

Adherence to these safety instructions is crucial for safe operation and to prevent personal injury or damage to the equipment.

  • Elektriohutus: Ensure the device is connected to a properly grounded power outlet with the specified voltage (220V, 60 Hz). Do not operate with damaged power cords.
  • Vedeliku käitlemine: Use only recommended heat transfer liquids. Avoid contact with skin and eyes. Refer to the safety data sheet (SDS) for the specific fluid used.
  • Ventilatsioon: Ensure adequate ventilation around the circulator to prevent overheating. Do not block air inlets or outlets.
  • Paigutus: Place the unit on a stable, level surface. Maintain sufficient clearance from walls and other equipment.
  • Äärmuslikud temperatuurid: Be aware of extreme temperatures of the bath fluid and external connections. Use appropriate personal protective equipment (PPE).
  • Hädaseiskamine: Familiarize yourself with the emergency stop procedure and the location of the main power switch.

3. Seadistamine

  1. Lahtipakkimine: Carefully remove the circulator from its packaging. Inspect for any shipping damage.
  2. Paigutus: Position the unit on a stable, level surface, ensuring at least 10 cm (4 inches) of clearance on all sides for proper ventilation.
  3. Vedeliku täitmine:
    • Open the bath cover.
    • Fill the bath with the appropriate heat transfer liquid up to the marked fill line. Refer to the fluid manufacturer's guidelines for compatibility and temperature range.
    • Veenduge, et pump oleks täielikult vee all.
  4. External Connections (if applicable): Connect external applications using appropriate tubing and fittings. Ensure all connections are secure to prevent leaks.
  5. Toiteühendus: Connect the power cord to the circulator and then to a grounded 220V, 60 Hz power outlet.
Three LAUDA Alpha cooling circulators, showing the front panel and bath opening.

Figure 1: LAUDA Alpha Cooling Circulators. This image displays three units, highlighting their compact design and front control panels.

4. Kasutusjuhised

  1. Toide sisse: Press the main power switch, usually located on the rear or side of the unit. The LED display will illuminate.
  2. Temperatuuri seadistamine:
    • Use the 3-button operation panel to navigate the menu.
    • Press the 'Set' button (or equivalent) to enter temperature setting mode.
    • Use the up/down arrow buttons to adjust the desired temperature.
    • Confirm the setting by pressing 'Set' again or waiting for the display to revert.
  3. Starting Circulation/Cooling: Once the temperature is set, the unit will automatically begin to circulate and cool the bath fluid to the setpoint. The compressor will activate as needed.
  4. Voolukiiruse reguleerimine: The flow rate can be adjusted using inserts and fittings to match the size of the bath or external application. Refer to the specific accessories for detailed instructions.
  5. Jälgimine: The large LED display provides clear readings of the current bath temperature and setpoint.
  6. Toide välja: To turn off the unit, press the main power switch. For extended periods of non-use, disconnect the power cord from the outlet.
LAUDA cooling circulator with a digital display showing 25.0 degrees Celsius.

Figure 2: LAUDA Cooling Circulator with Digital Display. This image shows a single unit from an angle, highlighting the digital control panel and the bath opening.

5. Hooldus

Regular maintenance ensures optimal performance and extends the lifespan of your circulator.

  • Cleaning the Bath:
    • Periodically drain the bath fluid using the drain connection on the rear.
    • Clean the bath interior with a mild detergent and soft cloth. Rinse thoroughly with distilled water.
    • Refill with fresh heat transfer liquid.
  • Cleaning Air Inlet: The front cover can be easily removed for cleaning the cool air inlet. Use a soft brush or vacuum cleaner to remove dust and debris from the condenser fins. Ensure the unit is powered off and unplugged before cleaning.
  • Vedeliku asendamine: Replace the heat transfer fluid according to the fluid manufacturer's recommendations or if it appears discolored or contaminated.
  • Välised ühendused: Regularly check all tubing and connections for wear, cracks, or leaks. Replace as necessary.

6. Veaotsing

ProbleemVõimalik põhjusLahendus
Seade ei lülitu sisseNo power supply; Power cord disconnected; Main switch offCheck power outlet; Ensure power cord is securely connected; Turn on main switch
Temperatuur ei saavuta sättepunktiInsufficient fluid level; Blocked air inlet; Incorrect setpoint; Overload on cooling capacityCheck and refill fluid; Clean air inlet/condenser; Verify setpoint; Reduce external load or check for proper sizing
Ekraanil veateadeSensor fault; Over-temperature; Low fluid levelRefer to specific error codes in the full technical manual (if available); Check fluid level; Allow unit to cool down
Ebatavaline müra või vibratsioonUnit not level; Loose components; Pump issueEnsure unit is on a stable, level surface; Inspect for loose parts; Contact service if pump noise persists

If troubleshooting steps do not resolve the issue, please contact LAUDA customer support.

7. Tehnilised andmed

  • Mudel: LCK 2907 Alpha RA 8
  • Operation Voltage: 220 V, 60 Hz
  • Temperatuurivahemik: -25 °C kuni 85 °C
  • Jahutusvõimsus: Up to 425 W (at specific temperatures, refer to full technical data sheet)
  • Juhtimine: 3-button operation with large LED display
  • Tühjendusühendus: Rear side for easy emptying
  • Sisaldab: Bath covers and pump kits
  • Tootja: LAUDA DR. R. WOBSER GMBH & CO. KG
  • Toote mudelinumber (sisemine): 1180A88EA

8. Garantii ja tugi

For warranty information, technical support, or service requests, please contact LAUDA customer service or your authorized distributor. Keep your purchase receipt and product serial number handy when contacting support.

Further documentation, including detailed technical specifications and safety data sheets for specific heat transfer fluids, may be available on the official LAUDA websaidil või nõudmisel.

Seotud dokumendid - LCK 2907 Alpha RA 8

Preview LAUDA Integral IN 590 XTW Process Thermostat Datasheet
Comprehensive technical data sheet for the LAUDA Integral IN 590 XTW Process Thermostat, detailing its features, specifications, performance data, and certifications.
Preview LAUDA Integral IN 590 XTW protsessitermostaat - tehniline andmeleht
LAUDA Integral IN 590 XTW protsessitermostaadi üksikasjalikud tehnilised andmed ja omadused, mis hõlmavad jõudlust, tööparameetreid, mõõtmeid ja jahutusvõimsust.
Preview LAUDA Universa Accessory Jet Pipe Assembly Instructions
Step-by-step assembly instructions for the LAUDA Universa accessory jet pipe, including safety guidelines, scope of delivery, and preparation steps. For LAUDA Universa MAX and PRO bath thermostats.
Preview LAUDA Integral IN 590 XTW Thermostat Process Technical Data Sheet
Detailed technical specifications and features of the LAUDA Integral IN 590 XTW Thermostat process, including operating ranges, power requirements, and connectivity options.
Preview LAUDA integreeritud protsessitermostaadid CO2 külmutusagensiga Kasutusjuhend
Kasutusjuhend LAUDA Integral IN 2040 XTW ja IN 2040 PW protsessitermostaatidele, mis kasutavad CO2 külmaainet. Hõlmab ohutust, seadistamist, kasutamist, hooldust ja tehnilisi andmeid.
Preview LAUDA Universa Bath Thermostats: Precision Temperature Control
Explore the LAUDA Universa series of bath thermostats, offering advanced temperature control solutions with modularity, AI-readiness, and sustainability. Discover ECO, PRO, and MAX models for diverse laboratory applications.